PROGRAMMABLE MODES FOR THE INSTRUMENTS

1
The key:
when all handpieces are laying, allows to choose the spray source (if the unit is fitted with the tank as an optional
item for physiological solution): yellow light (1) on means that the physiological solution is preset; no light on
means that pipeline water is preset.
Once a handpiece has been lifted, the key allows WATER and AIR selection of the spray of the lifted handpiece.
The YELLOW light lights up when WATER is preset, the GREEN light lights up when AIR is preset. If both lights
are off, then the handpiece will work in dry conditions; if you desire to program another condition, first press
the key so that the green light starts blinking: the blinking means that the microprocessor is waiting for the new
instructions.
By pressing once the key:
the yellow light will turn on; by pressing it again the green light will turn on; a further touch shall switch on both
yellow and green. Now the handpiece can work in the chosen conditions, but if it is placed back in the resting
array the microprocessor restores the previous conditions in memory. If on the contrary the new conditions
must be preserved, the key:
must be pressed: one blink shall confirm that the new instruction has been stored and the previous one
cancelled.
Similarly, by pressing the key:
first AIR, then WATER, and eventually both shall disappear. Therefore: yellow light on means that only water
is preset; green light on means that only spray air is preset; yellow and green lights both on at the same time
means that complete spray is preset; no light on means that dry operation is preset.
The violet key:
permits selection of WATER and AIR in the automatic purge at Foot-control release.
This function consists of a water or air or spray jet (emitted for an adjustable time) after the handpiece stops,
to the purpose of cleaning both the operating zone and the final parts of the hose. If you do not want to use
such device, just program the key with yellow and green lights off.
NOTE: the variables of this key (water and air) depend on the ones of the key - 1 - spray (see table A at page
42). When releasing the Foot-control lever, the preset automatic purge persists during a certain time; such time
can be adjusted according to the procedure explained in the section "PARAMETERS PRESETTING".
The violet key:
permits selection of WATER and AIR in the Foot-control c.b.. It is therefore possible to preset a normal c.b.
of sole air, or a sole water c.b., or eventually a spray c.b. without bur rotation. Introduction and change of modes
is obtained in the same way as seen for the SPRAY.
The violet key:
permits to adjust the maximum power provided to the instrument lifted (such max. power will be obtained when
the Foot-control lever is at end stroke). The power can be controlled and displayed in N% from 01% to 100%;
SPRAY
AUTO C.B.
1
C.B.
2

POWER ADJUSTMENT

3
ENTER