MODULES

The unit can hold 8 modules plus the Syringe one, at left and/or at right side of the modules holder. These eight modules
can be chosen among a wide range of types, and may include two or more modules of the same type (i.e. two or more
turbine modules).
TO REMOVE ANY MODULE FOLLOW THE PROCEDURE BELOW:
1
Switch off the unit and press the syringe levers up to the complete depressurization.
2
Press the release pin of the module (pin located under the module - see following paragraph "HANDPIECES
PRESSURE ADJUSTMENT"), holding the module or the group of modules following the one whose pin is pressed.
3
Once all air has flown out, shift the module or group of modules aside.
4
Press the disconnection pin of the module preceding the one to be taken out, and once all air is out slide the latter
aside.
5
Place back the module or group of modules previously taken off, or a new module in between, by sliding in the
supporting brackets. Taking care of proper matching between modules edges and tightness protection, switch the
unit on again.
6
If now the handpiece of the module placed in the position previously occupied by another is picked up, a beep is
heard and the display shows "AL04". In fact the microprocessor, whose memory matches that particular position
with the data of the previous module, is asking for the data relevant to the new one. To store them in, first push
the key:
to delete AL04 and push again to store the data shown on keyboard and display.
Such data can be varied according to the procedures above explained for the various functions.
If the position of other modules has been changed, this sequence of operations for reprogramming must be
repeated for all of them. All modules are fitted with flow rate adjustments for spray/c.b. water, spray/c.b. air and
drive/cooling air, provided they include such operative functions. The adjustment devices are located underneath
the module, beside the disconnection valve pin (see "HANDPIECES PRESSURE ADJUSTMENT").
ENTER
3
2
1
5
4
6
HANDPIECES PRESSURE ADJUSTMENT
1
Drive air (Turbines, Airmotor, Powder cleaner) or cooling
air (Micromotor, Light cure).
2
C.B. or spray air.
3
Spray water/C.B. water.
4
Whip balance adjustment.
5
Disconnection valve pin.
6
Syringe air adjustment.
43