Vega PLUS C 21 Instrucciones De Servicio página 51

Tabla de contenido

Publicidad

Command
aXWPOM<value1>!
aXRDU!
aXWDU<value1>!
aXRTU!
aXWTU<value1>!
aXRSR!
aXWSR<value1>!
aXRAPUR!
aXRPS!
aXRAPAM!
aXWAPPUL<value1>!
aXWAPPL<value1>!
aXWAPEC<value1>!
Read Power Operation Mode
Command
Response
aXRPOM!
a<value1><value2> <CR><lf>
Example:
Command
Response
0XRPOM!
0+0<CR><LF>
Write Power Operation Mode
Command
Response
aXWPOM <va-
a<value1><value2> <CR><LF>
lue1>!
VEGAPULS C 21 • SDI-12
Response
a<value1><value2><CR><LF>
a<value1><CR><LF>
a<value1><value2><CR><LF>
a<value1><CR><LF>
a<value1><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><CR><LF>
a<value1><value2><value3><va-
lue4><CR><LF>
a<value1><CR><LF>
a<value1><CR><LF>
a<value1><CR><LF>
a<value1><CR><LF>
Description
Write Power Operation Mode
Read distance unit
Write distance unit
Read temperature unit
Write temperature unit
Read stage reference
Write stage reference
Read access protection unlock result
Read parametrization state
Read access protection active mode
Write access protection parametrization lock
Write access protection parametrization unlock
Write access protection unlock with emergency co-
de
Description
a: sensor address
<value1>: power operation mode, +0 = low power mode,
+1 = normal power mode
<CR><LF>
Description
Description
a: sensor address
<value1>: power operation mode, +0 = low power mode,
+1 = normal power model
<value2>: VVO-Status eee
<CR><LF>
13 Anexo
51

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sdi-12Vegapuls c 21

Tabla de contenido