Vega PLUS C 21 Instrucciones De Servicio página 53

Tabla de contenido

Publicidad

Command
Response
0XWTU+0!
0+0+000<CR><LF>
0XWDU+4!
0+0+136<CR><LF>
Read stage reference
Command
Response
aXRSR!
a<value1><CR><LF>
Example:
Command
Response
aXRSR!
0+11.000<CR><LF>
Write stage reference
Command
Response
aXWSR <value1>! a<value1><value2> <CR><LF>
Example:
Command
Response
aXRSR+10.100!
0+10.100+000<CR><LF>
aXRSR+100!
0+10.100+134<CR><LF>
Read access protection unlock result
Command
Response
aXRAPUR!
a<value1><value2> <CR><LF>
Example:
VEGAPULS C 21 • SDI-12
Description
Valid data
No valid data
Current value is returned with a status 136
Description
a: sensor address
<value1>: stage reference +ss.sss [m], +ss.sss [ft], +dd-
ddd [mm], sss.ss [in]
<CR><LF>
Description
Description
a: sensor address
<value1>: stage reference +ss.sss [m], +ss.sss [ft], +dd-
ddd [mm], sss.ss [in]
<value2>: VVO-Status eee
<CR><LF>
Description
Valid data
No valid data is written
Description
a: sensor address
<value1>: result, +0 = success, +1 = failed, +2 = sequen-
ce error
<value2>: reason, +0 = without, +1 = wrong access co-
de, +2 =
<CR><LF>
13 Anexo
53

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sdi-12Vegapuls c 21

Tabla de contenido