Replacing The Spray Water Cartridge - Dentsply Sirona Midwest Stylus Plus Instrucciones De Uso

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6607290.book Seite 57 Montag, 5. September 2016 11:53 11
®
Operating Instructions Midwest Stylus
Plus Air-Driven-High-Speed Handpiece
A
B
6607290 D3673.201.01.01.02
➢ Remove the LED and re-insert it after rotating it 180° around its own
axis.
8.3.4

Replacing the spray water cartridge

Required accessories
● Spray water cartridge
®
T1 spray
● Midwest
Plus Aerosol Spray
If the spray water cartridge leaks or is clogged.
1. Remove the quick coupling from the supply hose.
Quick coupling
2. Remove the sealing washer (B).
3. Click the control ring into the maximum water flow. The spray water
cartridge (A) can be replaced only in this position.
C
4. Remove the spray water cartridge.
5. Spray oil lightly onto the new spray water cartridge.
6. Insert the new spray water cartridge.
8 Maintenance
57

Publicidad

Tabla de contenido
loading