Example
Determine the weight of a sample under extremely unstable ambient conditions
by calculating the average of 20 subweighing operations.
Settings (changes in the factory settings required for this example):
Menu: "Weigh averaging" application program, code 2.1.12
Step
1. Turn on balance, if power is off.
2. Tare balance
3. Select number of subweighing
operations
(20 measurements in this example)
4. Place sample on balance
(weight readout fluctuates;
here, for example, by about 275 g)
5. Start measurement
After 20 subweighing operations
If the print format is set to include
data ID codes, the following is printed:
6. Unload balance
7. Delete result
8. Repeat procedure starting from step 4, if desired.
54
Press key
Display/Printout
[ON/OFF]
[TARE]
+
+
[OK]
+
RES
+
(stable display)
[DELETE]
0.0 g
ref 20
(briefly)
8888
8888
20
19
...
1
275.5
g G
+
275.5 g
275.5
g G