17
4
5
If the process requires dilution to bring the measured sam-
ple within Kemio's measurement range, use a dilution factor.
If no dilution is required, the field can be left blank. If 1 or 0 is
entered for dilution, the label will display "- X" and no multi-
plication will be performed.
The displayed test result is calculated using the dilution fac-
tor programmed into the label.
For example:
A 100 mg/L sample needs diluting 4X. 1 part of sample
is diluted with 3 parts of deionized water. When a test is
performed using this label, the instrument will read 25 mg/L.
Using the dilution factor, 100 mg/L will be displayed on the
results screen.
Add a lower limit and/or upper
limit to set the range that is
considered a PASS for the test
application.
If a sample measured using
this label has a concentration
outside of this range, it will be
tagged as a FAIL.
6
Limits are applied to the calculated result. Press
SAVE to save the label in memory.
When configuring the label, if the test limits after dividing
by the dilution factor do not fall within the sensor method
test range, an error is displayed, and the label will not
close or save.
If multiple labels are set for a test type, the user must
select between them when performing a test. If a single
label is set, Kemio will automatically select it during a test.
If no labels are set, the label will be DEFAULT.
Edit a Label
Delete a Label
Select the label from the list.
Click on the number to be
changed. Enter a new number
and press SET or press X
to cancel.
Label names cannot be
edited. Create a new label
instead.
Select the label from the list.
Press the
button.
The label is deleted from
memory. Previous tests will
still be tagged with that label.
18