23
Check Standards
Kemio is supplied with check standards (KEMC10) to confirm
the instrument is working correctly.
Validating Kemio Using Check Standards
1
2
Start a check standard test by
A) pressing CONFIGURATION
then CHECK STANDARDS
or
B) pressing PERFORM TEST or
open the arm, and scan the QR
code on the outside of the check
standard box.
Open the arm, and insert the
first of four check standards
from the KEMC10 pack.
3
PASS
STANDARD S1
STANDARD S2
STANDARD S3
STANDARD S4
STANDARD S1
1
SUCCESSFUL
4
PASS
STANDARD S1
STANDARD S2
STANDARD S3
STANDARD S4
STANDARD S1
SUCCESSFUL
The check takes 10
seconds. If the instrument
is functioning correctly, a
PASS will display on screen
and the corresponding line
will turn green.
If Kemio displays a FAIL, clean
the check standard with a cloth
or alcohol wipe and repeat the
test with the same standard. If
the problem persists, contact
your Palintest representative.
Press NEXT or open the arm to
test the next standard. Repeat
with each standard in this way.
If all standards pass, Kemio will
display a summary report and
return to the home screen.
Results are stored in the data
log for traceability.
24