Or: by Ultrasound
• Use a low frequency ultrasound tank (25 to 50 kHz) and a detergent-
disinfectant product compatible with this process and with the MD,
D
• Friction of parts between them or with the tank can cause
appearance defects,
• Install light input/output protections on the body (Fig 7) to protect the
fronts of the fibre optic or the emitting surface of the LED,
• Immerse the components.
Rinsing and drying:
• Empty the spray pipe (Fig 4).
• Rinse well
,then dry each part.
5
VI.2.2 - Automatic preparation for sterilisation
e
• Only in a washing and disinfecting machine, appropriate for the
treatment of this type of MD. Follow the manufacturer's instructions
for the device.
• If the device is not equipped with pipes for contra angles / handpieces, then
disassemble the MD (see § V.2) and immobilise each part. Otherwise, put
the MD mounted on the pipe.
• The disinfection thermo cycle must be at least 10 minutes at 93°C (203°F).
x
• Check there is no residue and that all the parts of the MD are
completely dry at the end of the cycle.
• If the washing and disinfecting machine is near the place of use, it
D
is possible to also do step VI.1.
Re-assembly of MD:
Only assemble parts with the same serial number
• Attach the collar cartridge D to A head (only one direction possible) until it
completely plates (turn the cog until connection if required),
• Insert the head – collar cartridge unit in the G sleeve (connect a clean tool
in the head and turn it slightly so that the cog connects),
• Bring the F lever to the central position, until ratcheting. Make sure that
the depth stop system is in the right position.
E
Systematically check the condition of the front part by applying axial
traction.
VI.3 - Lubrication
Follow the spray lubricant manufacturer's instructions.
E
Wear appropriate protective clothing (splashes, etc.).
Keep away from all sources of heat or ignition.
In particular: do not smoke. Risk of inflammation.
5
Temperature < 38°C, distilled water is recommended instead of running water if this has
too much chlorine (cf. FD98-135 standard).
6
Exception : If the After-Sales Service has to replace a part, the serial number may be
different.
.
6
21