Step-by-Step Instructions
6.5

Cable Connection Test

A connection test feature is included in Propex Pixi
the cables:
1. Connect the measurement cable and turn on the device.
2. Connect the metal part of the connection hook to the lip clip. Make
sure that the accessories are cleaned properly before the test.
3. "Connection test" icon
see
Fig.
4. If no icon appears, the connection hook or the measurement cable
should be replaced.
Note
Measurement cable with attached lip clip and connection hook
constitute Applied Parts of the device.