INCUBATOR COMPONENTS

Remote alarm terminal

The alarm of this product can be transferred to a remote location by connecting the external alarm unit to
the remote alarm terminal. For the behaviour of remote alarm output, refer to pages 78–82.
The terminal of the remote alarm is provided at the
back upper right of the unit (see the right figure). The
alarm is output from this terminal. Contact capacity is
DC 30 V, 2 A.
Table 1 shows the behaviour of the remote alarm
when the BUZZER STOP key is pressed.
Notes:
• The door alarm does not work remotely. Refer to
pages 81 and 82.
• Wiring of the remote alarm should be performed by
qualified service personnel.
• It is recommended to use standard signal and interface
cables with a maximum length of 30 meters.
Table 1 Behaviour of remote alarm when pressing BUZZER STOP key
Remote Alarm setting
(Refer to pages 55-61)
ON: Remote alarm setting not
linked with BUZZER STOP key
OFF: Remote alarm setting
linked with BUZZER STOP key
1
In the case of Err01, Err11, Err12, and Err18, the condition returns to normal.
:
*
Use a twisted shield wire for the connection.
Type: UL 2343, UL 2448, UL 2464, UL 2552, UL 2623.
Length: 30 m max.
Connecting
Normal
terminal
condition
COM.-N.C.
Close
COM.-N.O.
Open
COM.-N.C.
Close
COM.-N.O.
Open
Abnormal condition
(Including during the power outage and when
power supply cord is disconnected)
When pressing the BUZZER
Open (still abnormal)
Open
Close (still abnormal)
Close
Open
Close (Return to normal)
Close
Open (Return to normal)
Remote alarm
terminal
STOP key
*1
*1
15