Medline MDS3001 Guía De Operación página 11

Monitor de presión arterial con manguito para brazo totalmente automático
Ocultar thumbs Ver también para MDS3001:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

automated sphygmomanometers.
9. Information regarding potential electromagnetic or other interference between the
blood pressure monitor and other devices together with advice regarding avoidance
of such interference please see part ELECTROMAGNETIC COMPATIBILITY
INFORMATION. It is suggested that the blood pressure monitor be kept at least
30 cm away from other wireless devices, such as WLAN unit, microwave oven,
etc. It can't be used near active HF SURGICAL EQUIPMENT and the RF shielded
room of an ME SYSTEM for magneticresonance imaging, where the intensity of
EM DISTURBANCES is high.
10. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the
procedure of blood pressure measurement, a signal of
Under this condition, the Electronic Sphygmomanometer can continue functioning,
but the results may not be accurate; it's suggested that you consult with your
physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
a) The coefficient of variation (CV) of pulse period >25%.
b) The difference of adjacent pulse period ≥0.14s, and the number of such pulse
takes more than 53 percent of the total number of pulse.
-11-
 
will be displayed.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mds3001plusMds3001u

Tabla de contenido