64 | WBE 4120 / WBE 4120 D / WBE 4120 DT | Diagnostics
10.5.2
Diagnosis width potentiometer(IN3)
Difetto
The value remains blocked
at scale end or between 0
and 40
The value remains blocked
between 100 and 16000.
10.5.3
Diagnosis diameter potentiometer (IN4)
Difetto
The value remains blocked
at scale end or between 0
and 40.
The value remains blocked
between 100 and 16000.
The values passes from 0 du-
ring the run..
!
After having replaced the potentiometer or after
having detected that the fastening screw is loose,
before proceeding with calibration, follow the inst-
ructions reported here.
|
1 695 600 579
2016-08-02
Correzione
+
Check that the cable is cor-
rectly connected.
+
Check that the ring nut is
correctly fixed.
+
Replace the connection
cable.
+
Replace the potentiometer.
+
Check that the ring nut is
correctly fixed.
+
Replace the connection
cable.
+
Replace the potentiometer.
Correzione
+
Check that the potentiome-
ter is correctly fixed to the
measurement arm.
+
Replace the connection
cable.
+
Replace the potentiometer.
Check that the potentiome-
+
ter is correctly fixed to the
measurement arm.
+
Replace the connection
cable.
+
Replace the potentiometer.
Re-position the potentiometer
(see following Instructions).
10.6
Ingressi ed uscite digitali
10.6.1
Protection micro switch
i
When the cover is lowered, if the machine functions
correctly, the ON sign appears.
Should the micro show problems, check the position
of the functioning cam and of the same micro, as
well as the cables' connection (see electric layout).
!
A malfunctioning of the micro can cause the pneu-
matic flange to block.