  • MEXICANO, página 129
Indicator light/
audio:
Status/Indicator:
Solid Yellow
The audible alarm
will sound
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
can be paused
by pressing the
Audio Pause
button. When
pressed the light
above the button
will illuminate.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Impact to therapy:
Device will continue
to operate normally
for 3 hours and will
stop operating once
the battery is fully
discharged.
Device will continue
to operate normally
for 1 hour and will
stop operating once
the battery is fully
discharged.
27
Steps to remedy:
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
EN