Indicator light/
audio:
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
Impact to therapy:
The battery has
only 2 minutes
of therapy times
remaining. Device
must be plugged
into electrical outlet
to continue therapy.
If it is not plugged
into an electrical
outlet it will stop
operating once
the battery is fully
discharged.
Device must be
plugged into an
electrical outlet to
continue therapy.
Device must be
plugged into an
electrical outlet to
continue therapy.
If it is not plugged
into an electrical
outlet it will stop
operating once
the battery is fully
discharged.
29
Steps to remedy:
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Contact your healthcare
provider if the use of a battery
is preferred.