Indicator light/
audio:
Status/Alarm
Indicator:
Solid Yellow
The audible alarm
will sound.
The audible alarm
can be paused
by pressing the
Audio Pause
button. When
pressed the light
above the button
will illuminate.
Impact to therapy:
Device will continue
to operate, but may
not provide the
prescribed therapy.
33
Steps to remedy:
• Press down around the
edges of your dressing. If
your dressing begins to
recompress, you have
resealed your dressing and
addressed the air leak.
• Check the connectors on
the tubing between your
dressing and the canister.
Ensure the connection is
secure (see images on pages
20 and 22.)
• Disconnect the quick click
connectors and close the
tethered caps to seal the
tubing. If the alarm continues,
replace the canister.
• If you cannot locate the
source of your air leak, contact
your healthcare provider.