STEP 4: Check if the Test Result is in Range.
In less than 7 seconds, the control
solution test result appears on the display.
Compare the test result with the range
printed on the test strip vial. The test
result should fall within the printed range.
See Figure 22.
Voice Speaks: "Blood glucose
is 59 mg/dL."
Out of Range Results
If test results fall outside the range printed
on the test strip vial, check "Trouble-Shooting
Guide" located in the "System Troubleshooting"
Section and repeat the test.
If you continue to get out-of-range results, it means that the
system or the control solution may not be working properly.
DO NOT use the system to test your blood glucose level.
If you are unable to resolve the problem, contact Technical
Support at 1.800.243.2636.