Descargar Imprimir esta página

Calibration; Ckt Calibration - Adam Cruiser CKT Serie Manual Del Usuario

Publicidad

Idiomas disponibles

Idiomas disponibles

uSB
uSB menu
S-id
Set Scale ID
U-id
Set User ID
rECHAr
Indicates
battery chartge
6.0

CALIBRATION

OIML TYPE APPROVAL: For the CKT-M models, the calibration is locked either by a sealed
jumper on the underside of the scale, or by a calibration count on the display. If the seal is
broken or tampered with, the scale needs to be re-verified by an authorised certification
body and re-sealed, before it is used legally. Contact your local metrology standards office
for further assistance.
6.1

CKT CALIBRATION

You need to enter a secure menu by entering a passcode when requested.
• Press [Tare] ↵ once, during the initial counting of the display after the power is
turned on.
• The display will show "P" requesting for the passcode number.
• The fixed passcode is "1000"
• Press the [Tare] ↵ key
• The display will show "u-CAL"
• Press the [Tare] ↵ key and the display will show "noload" to request all weight be
removed from the platform.
• Press the [Tare] ↵ key to set the zero point
• The display will then show the calibration weight suggested in the display. If the
calibration weight is different from the value shown, Press [CE] to clear the current
© Adam Equipment Company 2019
AC oFF –for selecting the option of
accumulating manually or turned off (AC OFF
/ AC ON)
Manual –selecting by output
ATP – printer type ( ATP/LP 50)
Copy 1 : select the number of copies (1-8)
Comp : many lines or Sinp: simple – one line
LF/CR – line feed and carriage return to feed
printer paper (0 -9 lines)
PC options:
4800 – for setting the baud rate – use digits
[1] and [6] to choose from the options:
1200/2400/4800/9600/38400/57600/115200.
Adam – for connecting to Adam DU software
(use digits [1] and [6] to select between 'cbk'
or 'nbl' option)
int (interval) – select the interval per second
for sending data to a PC ( 0, 0.5, 1, 2, 2.5, 3,
3.5, 4, 4.5, 5, 5.5, 6)
PC- same as per rs 232
Print – same as per rs232
To be entered manually
To be entered manually
the
Without adaptor –shows battery voltage
With the adaptor shows charging current (mA)
16
Manual
ATP
Copy 1
Comp
1 LFCr
4800
Int 0
000000
000000
-

Publicidad

loading