value then enter the correct value as an integer value, it is not possible to have
fractions of a kilogram or pound. For Example:
20kg =
Press [Tare] ↵ to accept the calibration value and the display will now show "Load".
Place the calibration weight on the platform and allow the scale to stabilise as
indicated by the stable indicator.
Press [Tare] ↵ to Calibrate
When calibration is done the scale will restart and return to normal weighing.
After calibration, the scale should be checked whether the calibration is correct. If
Necessary, repeat calibration.
7.0

BATTERY

Battery Life
Battery life is approximately 90 hours without mains connection.
Charging
The battery symbol is always shown up on the display. When the battery needs to be
charged, the symbol on the display is shown empty and the backlight will be disabled to
save power.
The scale will still operate for about 10 hours after which it will automatically switch off to
protect the battery.
Plug into the mains power supply and charge for 12 hours for full capacity. The scale does
not need to be turned on.
8.0

MODULE OPTION COVERS/ BOARDS

Module option boards are located underneath the scale with covers to protect the ports.
© Adam Equipment Company 2019
Module cover boards