6 Maintenance
6.1 Servicing and Inspection
The product is maintenance-free, provided the cleaning and
safety instructions outlined below are adhered to.
To ensure good performance and safety the following checks
should regularly be carried out:
•
Check the product and accessories regularly for damage.
Do not use the product if defective.
•
In case of damage or malfunction it is recommended to
contact the provider.
•
Ensure that the product is securely assembled and that
screws are properly tightened.
•
If the product is not used for a prolonged period,
charge the battery completely and then raise and lower
the seat frame 3 times without load. Repeat this
process every 3 months. This helps to maintain the
service life of the battery.
WARNING!
Risk of Injury and Damage
If the product is improperly adjusted it can cause
accidents and severe injuries.
– The product must be repaired by a qualified
technician.
– Damaged parts have to be replaced by a
qualified technician.
– Only use original Invacare spare parts.
26
WARNING!
Risk of Injury or Damage
Incorrectly fitted castors or drives can lead to
accidents.
– The castors and drives must be replaced by a
qualified technician.
6.2 Cleaning and Disinfection
6.2.1 General safety information
CAUTION!
Risk of contamination
– Take precautions for yourself and use
appropriate protective equipment.
CAUTION!
Risk of electric shock and product damage
– Switch off the device and disconnect from
mains, if applicable.
– When cleaning electronic components consider
their protection class regarding water ingress.
– Make sure that no water splashes to the plug
or the wall outlet.
– Do not touch the power socket with wet hands.
1556395-D