General Information

Refer also to installation and operating instructions for
VistaScan ....
Symbols
EN
0297
CE labelling
Batch number
Manufacturer
Intended purpose
The image plates are used to create intraoral and extraoral X-ray
images.
Intended use
The image plates are designed for use solely with Dürr Dental
image plate scanners and Dürr Dental light protection covers in
dental applications.
CAUTION
Image plates are toxic
Image plates that are not packed in a light protection
cover can lead to poisoning when placed in the mouth
or swallowed.
Only place image plates in the patient's mouth in a
i
light protection cover.
Do not swallow the image plate or parts of it.
i
If the image plate or parts of it have been swallowed,
i
consult a specialist doctor immediately and remove
the image plate.
If the light protection cover has been damaged in the
i
patient's mouth, rinse the mouth thoroughly with lots
of water. Do not swallow the water in the process.
Image plates are flexible like X-ray film. However, the image
i
plates should not be bent.
Do not scratch the image plates. Do not subject the image
i
plates to pressure from hard or pointed objects.
Protect the image plates against sunlight and ultraviolet light.
i
Store image plates in a light protection cover or plate cassette
of the correct size.
Image plates will be pre-exposed on exposure to natural radi-
i
ation and stray x-ray radiation. Protect erased and exposed
image plates from X-ray interference.
If the image plate has been stored for longer than one week,
erase the image plate prior to use.
Do not store image plates under hot or moist conditions.
i
When used properly, image plates can be exposed, read and
i
erased several hundred times provided there is no mechanical
damage.
Replace the image plate if there are any signs of damage (e.g.
protective layer is damaged or visible scratches) that could
interfere with the diagnosis.
Image plates that have a production or packaging defect will
i
be replaced by Dürr Dental in the same quantity. Claims can
only be accepted within 7 working days after receipt of the
goods.
Classification
Medical device directive (93/42/EEC) ............................. Class IIa
Storage and transport
Temperature (C°) ................................................................. < 33
Relative humidity (%) ........................................................... < 80
Usage
Temperature (C°) .............................................................18 - 45
Relative humidity (%) ........................................................... < 80