7.6 Neutral electrode settings and attachment

In the monopolar operating mode, a neutral electrode is required.
Do not use undivided but only separate neutral electrodes.
• Place the neutral electrode on the patient.
• Connect the cable of the neutral electrode to the electrode.
• Connect the cable of the neutral electrode to the socket (3) on the front of the device.
After attaching the neutral electrode to the patient, the HF unit recognizes the neutral
electrode automatically (cf. illustration on the left and chapter 4.2
onnte nicht gefunden werden.
geteilten und ungeteilten Kontaktflächen.
!
If the displayed neutral electrode type does not agree with the used
neutral electrode type, this is indicative of poor adhesion of the neutral
electrode on the patient, a defective neutral electrode or a defective cable!
). Das HRF 100 x-touch erkennt Neutralelektroden mit
Please note the following information when attaching the
neutral electrode