9 Protective measures and warnings

Please observe also the instructions provided on working with the neutral electrode in chapter
7.6 the operating instructions with the accessories and the applicable safety measures!
!
Before any procedure, check the cables, accessories (handles,
defective insulation. These can result in burns if they come into
Before any procedure, check the cables, accessories (handles, electrodes, forceps) and
foot switches for any signs of visible irregularity. Do not use cables or instruments with
broken or defective insulation. These can result in burns if they come into contact with the
!
The device must not be used in areas where there is a risk of explosion.
!
The device may only be repaired by HEBU or a representative authorized by HEBU.
!
It must be ensured that the device is only operated by people who are familiar with the
operating instructions.
!
The device, including the connecting cables, should be checked before each use.
!
Faults must be eliminated in good time
!
All plug-in connections may only be carried out with compatible accessories.
!
Use the greatest care when working with the neutral conductor!
!
Protect the patient from contact with earthed metallic elements (OP table, infusion stand
etc.) during use.
!
Avoid contact between supply cables and patients, other cables or instruments! Ensure
that there are no loops in the cables of handles and electrodes or fixture to the OP table.
!
Avoid small-area skin-to-skin contact in the patient (e.g. a finger making contact with the
thigh).
!
Avoid contact between HF instruments and non-insulated instruments..
electrodes, forceps) and foot switches for any signs of visible
irregularity. Do not use cables or instruments with broken or
contact with the skin!
skin!