HBS x-touch 200i
Areas for attachment of the neutral electrode
◼
Attach the neutral electrode to clean, dry skin.
◼
The electrode must be attached close to the operation area, but with a minimum direct distance
from the operating field of 20 cm.
◼
Attach over tissue with a good supply of blood without dips or curves in the skin, for instance on
the upper arm or thigh. (cf. picture below)
◼
Do not attach the neutral electrode to projecting surfaces.
◼
Do not attach above bones, scars, cuts or scratches.
◼
The electrode must not be attached to surfaces with a high amount of hair growth. If necessary,
shave the relevant area before attaching.
◼
Do not attach above particularly adipose areas such as the abdomen or buttocks.
◼
The neutral electrode must not be attached above implants.
◼
Electrodes of monitoring systems must be placed at a distance of at least 20 cm from the operation
area and from the position of the attached neutral electrode
◼
Correct attachment of the neutral electrode on adult patients
◼
Correct attachment of the neutral electrode on children
◼
Attachment on patients with pacemakers
Note! Consult the treating cardiologist!
RIGHT
WRONG!
EN
49