Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be
solved, contact your equipment supplier or ResMed. Do not attempt to open the
device.
Problem
No display.
Insufficient air
delivered from
VPAP Auto
VPAP Auto does not
start when you
breathe into the
mask.
24
Possible Cause
Power not connected.
Ramp or settling time is in use.
Air filter is dirty.
Air tubing is kinked or
punctured.
Air tubing not connected
properly.
Mask and headgear not
positioned correctly.
Cap(s) missing from access
port(s) on mask.
Pressure required for treatment
may have changed.
Power cord not connected
properly.
Power outlet may be faulty.
Breath is not deep enough to
trigger SmartStart.
Solution
Ensure the appropriate power
cord is connected to your VPAP
Auto and the power switch (if
available) is on.
Wait for air pressure to build up
or change ramp or settling time.
Replace air filter.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and
headgear.
Replace cap(s).
See your clinician to adjust the
pressure.
Connect power cord firmly at
both ends.
Try another power outlet.
Take a deep breath in and out
through the mask.