Sartorius PMA.Vision Instrucciones De Instalación página 93

Balanzas para la mezcla de pinturas en áreas potencialmente explosivas de la zona 1
Ocultar thumbs Ver también para PMA.Vision:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Hazardous Location
Class I, Division 1, Groups C,D, T4
Class I, Zone 1, Group IIB, T4
SARTORIUS
Model
VIS1X.....-.....
1)
USA: The installation must be in accordance with the National Electrical Code
ANSI / ISA-RP 12.6.
Canada: The installation must be in accordance with the Canadian Electrical Code
2)
No connection to any device that uses or generates in excess of 250Vrms or 250Vdc.
3)
USA: The apparatus must be connected to a suitable ground electrode per National Electrical Code
Article 504 or 505. The resistance of the ground pad must be less than 1 ohm.
Canada: The apparatus must be connected to a suitable ground electrode per Canadian Electrical Code
18. The resistance of the ground pad must be less than 1 ohm.
4)
range: 0°C .... +40°C (+32°F .... + 104°F)
5)
WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
If your equipment needs to be repaired, use only original spare parts supplied by the manufacturer!
6)
Only the Sartorius cable type YCC01-0052Mxx (XX = 10 for 10m, 20 for 20m; 30 for 30m) may be used. Maximum
cable length: 100ft (30.5m). The equipment operator is responsible for any non-Sartorius cables used.
7)
Be sure to follow the installation, operating, maintenance and servicing instructions given in the manuals
supplied.
8)
The equipment shall be installed in such a way that it is protected against the entry of solid foreign objects or
water capable of impairing the safety of the apparatus. Reduce the risk of mechanical damage to a minimum.
9)
Exposure to UV radiation is not allowed!
10) The connecting cable of the display unit must be prevented against damage and stress caused by strain.
11) Prior to opening the equipment, disconnect the power supply or make sure that there is no potentially explosive
atmosphere or any other explosion hazard in the surrounding area!
12) The data cable connected to the scale (weighing unit) is considered as intrinsically safe circuit. Check the correct
function of the data transfer before you use the equipment in a hazardous location.
13) If the equipment does not operate properly, unplug it immediately from line power (mains supply)!
14) Avoid generating static electricity. Use only a damp cloth to wipe down the equipment. The equipment operator
shall be responsible for preventing any risks caused by static electricity.
15) Keep chemicals and other agents, which can corrode the housing seals and cable sheaths, away from the
equipment. These agents include oil, grease, benzene, acetone and ozone. If you are not sure about the safety of a
certain substance, please contact the manufacturer.
16) Use equipment only in the temperature ranges indicated. Avoid exposing the equipment to heat.
17) At reasonable intervals, have your equipment installation checked for proper functioning and safety by a trained
and certified technician.
18) Any tampering with the equipment by anyone, other than repair work done by authorized Sartorius service
technicians, will result in the loss of EX conformity and in the forfeiture of all claims under the manufacturer's
warranty. Only authorized specialists may open the equipment.
19) Modifications, including those to be carried out by Sartorius employees, may be permitted only after the express
written authorization has been obtained from Sartorius.
2015-02-20
Dr. D. Klausgrete
Sartorius Cable Type
6)
YCC01-0052Mxx
3)
Appendix
Non-Hazardous Location
2) Ethernet
SARTORIUS
Model
YCO16-Z
2) Power supply
3)
®
, NFPA 70, Article 504 or 505 and
®
, Section 18.
Installation Instructions VIS1X
(optional); e.g.
Sartorius Model
YPS07-USB
®
, NFPA 70,
®
, Section
Sheet
of
93

Publicidad

Tabla de contenido
loading

Tabla de contenido