Transport / shipment
Caution:
The drive unit and probe can be dam-
aged during transport or shipping.
The following instructions must be ob-
served when transporting or shipping
the instrument. Otherwise, all warranty
claims will be void!
When transporting internally in the carry case:
1.
Place the red shipping clamp from below
on the vee-block holder (9) (see information
sheet 3763316).
2.
Switch off the MarSurf PS 10 and stow it in
the carry case.
When shipping:
1.
Switch off the MarSurf PS 10.
2.
Remove the drive unit from the MarSurf
PS 10 (see in Section 4).
3.
Place the red shipping clamp from below
on the vee-block holder (9) (see information
sheet 3763316).
4.
Stow the drive unit in the dedicated com-
partment of the corresponding shipping
box.
5.
Pack the MarSurf PS 10 (in its carry case)
and the rest of the accessories into the ship-
ping box.
18
6.1
Checking the integrated standard
To determine the Rz value of the integrated
standard:
1.
Position a suitable roughness measuring
instru ment
in such a way that the probe
9
is on top of the standard and a traversing
length of 5.6 mm (0.220 in) can be meas-
ured (see figure 3).
2.
Set the traversing length Lt = 5.6 mm
(0.220 in) on the measuring instrument.
3.
Perform one or several measurements and
read off the measured Rz value in each case.
4.
Average the measured Rz values.
5.
Enter the calculated Rz value as the Rz
nominal value during the next calibration
with the MarSurf PS 10.
6.2
Replacing the battery
The maximum capacity of the built-in battery
gradually decreases over a long period of use
and, therefore, may need to be replaced.
The battery should only be replaced by
trained specialist staff.
Only the lithium-ion battery of 3.7 V /
3150 mAh (3030057) may be used,
other wise the instrument might be de-
stroyed.
To replace the battery (see fig. 6 in back flap):
1.
Use a Phillips screwdriver to undo and re-
move the four screws (25) on the bottom
of the MarSurf PS 10.
2.
Set the MarSurf PS 10 on its vee-block base.
9
e.g. MarSurf SD 26 drive unit with BFW-250 skidless
probe and MarSurf M 400 or MarSurf XR 1 evaluation
instrument
Mahr GmbH, MarSurf PS 10