INSTRUCTIONS FOR USE
d. To prevent patient and/or user injury and/or equipment damage, verify the
device attaching clamps completely touch the table-side rails and are firmly in
place. Test the locking mechanism to ensure no movement when elevated or
pushed.
e. Product will be damaged, if it is cleaned with bleach.
f. Prior to using this or any other type of medical apparatus with a patient, it is highly
recommended that you read the Instructions for Use and familiarize yourself with
the product.
g. Do not use if the product shows visible damage.
h. Read and understand all warnings in this manual and on the device itself prior to
use for patient.
i.
Device function should be checked prior to each usage.
j.
This device should only be operated by medically trained personnel.
k. Additional positioning devices should be used when using the stirrup in
Trendelenburg or reverse Trendelenburg.
l.
The patient's toes may extend beyond the end of the stirrup boot.
m. The product with specified rail clamps.
n. Check for patient contact pressure points and consult the physician before use.
o. Warn the user against using non-approved padding.
p. To secure the patient's legs in the boots prior to adjusting.
CAUTION:
a. Do not exceed safe working load shown in the product specification table.
b.
Do not clean stirrup by immersing it in liquid or using high-pressure spray. Such
action may damage product .
c. Do not raise the thigh closer to torso than 90º as patient may experience nerve
damage.
d. To prevent patient and/or user injury and/or equipment damage: all
modifications, upgrades, or repairs must be performed by an Allen authorized
specialist. Failure to comply may void the warranty.
e. Keep patient and caregivers away from the area between the gas spring and
the main support tube.
lithotomy.
Document Number: 80028203
Version: A
Possible pinch point hazard when in extreme high
Page 17
Issue Date: 18 MAR 2020
Ref Blank Template: 80025118 Ver. E