CLEANING AND MAINTENANCE

IMPORTANT:
For this operation, use all necessary protective measures (gloves, mask, overalls etc.)
then destroy contaminated consumables (antibacterial cartridges, condensates etc.) in
according with that country's applicable laws (regulation incinerator).
Maintenance is to be entrusted to trained and approved personnel by the exploiting, with
all necessary knowledge and acquainted with best practice.
Not to carry out maintenance operations when the device is in use.
If the spare parts used during maintenance operations are not corresponding to the
manufacturers' requirements, the manufacturer declines all responsibility in the event of
an incident.
The modification of the central units is forbidden.
Modification of the device is forbidden without the approval of the manufacturer.
If the central is modified following manufacturer agreement, a control and a suitable test
must be realized to make sure that the central is always safe to use.
To avoid risk of damage or prevent equipment become dangerous is essential use the compatible
components provide by the manufacturer.

FILTRATION

bleed PYREX vessel
change anti-bacteria filter(s)
MAIN and SECONDARY RECEIVERS
general inspection of condition
call in an outside inspection body
ELECTRICAL BOX
clean and dust using a soft brush
check the connections in the cabinet(s)
check the connections at motor terminals
operate the mechanical back-up systems
check the intensity of the motors
VACUUM PUMPS
VACUUM HOSE
Check general condition
Change
NON RETURN VALVE
Check tightness
517463-07
HOSPIVAC G E350.R
Monthly
Quarterly
Annually
2 years
5 years
Regular inspection
According to legislation
Refer to the start-up and maintenance manual
E600.R
DN80 ISOLATING VALVE
Check correct working order
FLEXO SOLENOID VALVE
Check correct working order
Changement de son filtre
PRESSURE INDICATOR (electronic sensor, vacuum gauge)
Check measures for correctness
NETWORK ALARM
Check correct working order
EXHAUST HOSE
Check general condition
Change
DELIVERY POT
Drain if necessary
11.1. Filtration
IMPORTANT:
For this operation, use all necessary protective measures (gloves, mask, overalls etc.)
then destroy contaminated consumables (antibacterial cartridges, condensates etc.) in
according with that country's applicable laws (regulation incinerator).
During and after maintenance, respect basic hygiene rules.
Before ordering tools used for maintenance, clean them with a disinfectant.
11.1.1.
Bleed the PYREX vessel
 Close upstream and downstream valves,
 Bring to atmospheric pressure by pressing the red button on the filter nose,
 Unscrew the vessel, empty it and screw it back into position.
 Put the upstream and downstream valves back into working position.
11.1.2.
Changing the anti-bacteria filter
Systematically replace the anti-bacteria cartridge every 12 MONTHS or more frequently if pressure on the
network drops too low.
 Isolate the filter by closing its upstream and downstream valves,
 Bring to atmospheric pressure by pressing the red button on the filter nose,
 Unscrew the chamber, remove the filter element.
 Fit the new filter element and clean the chamber.
 Then, once reassembled, position the upstream and downstream valves back to their working positions.
Monthly
Quarterly
Annually
2 years
5 years
Regular inspection