Guidance and manufacturer's declaration - electromagnetic emissions
The electrical breast pump NUK e-Motion is intended for use in the electromagnetic environment specified
bellow. The costumer or the user of the electrical breast pump NUK e-Motion should assure that it is used in
such an environment.
Immunity test
IEC 60601-1 test level
Electrostatic
+/- 6kV contact
discharge (ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast
+/- 2 kV for power supply
transient/ burst
lines
IEC 61000-4-4
+/- 1 kV for input/ output
lines
Surge
+/- 1 kV differential mode
IEC 61000-4-5
+/- 2 kV common mode
Voltage dips, short
<5% for 5s
interruptions and
<5% for 0,5 cycles
voltage variations on
40% for 5 cycles
power supply input
70% for 25 cycles
lines
IEC 61000-4-11
Power frequency
3 A/m
(50/60Hz) magnetic
fields
IEC 61000-4-8
Compliance level
+/- 6kV contact
+/- 8 kV air
+/- 2 kV for power supply
lines
+/- 1 kV for input/ output
lines
+/- 1 kV differential mode
+/- 2 kV common mode
<5% for 5s
<5% for 0,5 cycles
40% for 5 cycles
70% for 25 cycles
3 A/m
111
Electromagnetic environ-
ment - guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
materials, the relative humidity
should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the electrical breast
pump NUK e-Motion requires
continued operation during
power mains interruption, it is
recommended that the electri-
cal breast pump NUK e-Motion
be powered from an uninter-
ruptible power supply or bat-
tery.
Power frequency magnetic
fields should be at levels charac-
teristic of the typical location in
a typical commercial or hospital
environment.