Guidance and Manufacturer's Declaration: Electromagnetic Immunity
SERFAS Energy Systemis intended for use in the electromagnetic environment specified below. The customer or
the user of SERFAS Energy Systemshould ensure that it is used in such an environment.
Immunity
Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the SERFAS Energy
Systemsystem is used exceeds the applicable RF compliance level above, the SERFAS Energy Systemsystem should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the SERFAS Energy Systemunit.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IEC 60601
Compliance
Test Level
3 Vrms
150 kHz to
80 MHz
3 V/m
80MHz to
2.5 GHz
and reflection from structures, objects, and people.
Electromagnetic Environment: Guidance
Level
Portable and mobile RF communications
equipment should be used no closer to any
part of the SERFAS Energy Systemsystem,
including its cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
3 V
Recommended Separation Distance
3 V/m
d
d
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
compliance level in each frequency
Interference may occur in the vicinity of
equipment marked with the following
transmitter.
d
=
1.17 P
=
1.17 P
80 MHz to 800 MHz
=
2.33 P
800 MHz to 2.5 GHz
distance in meters (m).
(a)
, should be less than the
(b)
range
.
symbol:
EN-31