DENTSPLY Sani-Tip L-3 Manual Del Usuario página 6

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 14
cleaning: automated
None.
cleaning: manual
1. Clean all external surfaces of the Sani-Tip
a mild detergent and water solution. Dry with a lint-free cloth.
2. Remove the conversion kit from the air/water syringe.
3. Clean all internal surfaces of the adapter, thumb nut, plastic cone and o-ring with a mild
detergent and water solution. Dry with a lint-free cloth.
4. Further process the adapter and thumb nut following the Sterilization procedure listed.
5. The plastic cone and o-ring are not steam autoclave sterilizable. Further process following
the disinfection procedures listed.
disinfection:
Disinfect the plastic cone and o-ring using standard disinfection solutions.
Packaging:
Place thumb nut and adapter in a paper or paper/plastic steam sterilization pouch. Components may
be loaded into dedicated instruments tray.
sterilization:
Use a steam autoclave. Place bagged adapter and thumb nut into the steam autoclave, paper side up
when using a paper/plastic pouch. After warm-up is completed, operate at a sterilizing temperature and
pressure of 134ºC/273ºF to 137ºC/279ºF and pressure of 216kPa/31psi for 3 to 12 minutes.
Alternate method: Place non-bagged adapter and thumb nut into the steam autoclave. After warm-up is
completed, operate at a sterilizing temperature and pressure of 134ºC/273ºF to 137ºC/279ºF and pres-
sure of 216kPa /31psi for 3 to 4 minutes. Components should be used immediately.
drying:
To dry, use the drying cycle of the autoclave. Set cycle for 20-30 minutes.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
• Check for distortion, damage and wear. Discard damaged, worn or corroded components.
storage:
To maintain sterility, instruments should remain bagged until ready to use.
ManUfactUrer contact
In the United States, contact DENTSPLY Professional Customer Service Technical Support at
800-989-8826. For areas outside the United States, contact your local DENTSPLY Representative.
6
conversion kit, including the air/water syringe with
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

122326n5

Tabla de contenido