Care And Handling; Manual Disinfection And Cleaning; Mechanical Cleaning And Disinfecting; Care And Inspection - B.Braun BA720R Instrucciones De Manejo/Descripción Técnica

Dermatomo para injertos de malla
Ocultar thumbs Ver también para BA720R:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Aesculap Power Systems
Skin mesh dermatome BA720R

5. Care and handling

Note
For patients with Creutzfeldt-Jakob Disease (CJD), suspected CJD or
possible variants of CJD, the relevant national regulations concerning the
reprocessing of the products must be observed.

5.1 Manual disinfection and cleaning

! Wipe off the product with a clean, lint-free cloth soaked in cleaning
and disinfecting agent.
- or -
In order to prevent formation of sticky or encrusted residues, spray the
product with a disinfectant spray.
! Wipe off the component surfaces with a clean, lint-free cloth.
! Follow the disinfectant manufacturer's instructions.
! To avoid damage to cutting cylinder 2, clean the cutting cylinder with
a soft brush.

5.2 Mechanical cleaning and disinfecting

Damage to the skin mesh dermatome caused by
incorrect mechanical cleaning!
! Only clean the skin mesh dermatome in a suitable
CAUTION
storage aid.
! Always use a suitable cleaning and disinfection
program for mechanical cleaning.
! Do not use alkaline cleaning agents.
! Carry out mechanical cleaning immediately after the operation.
! Install storage aid GB688R in a suitable wire basket (e.g. JF214R).
! Insert the skin mesh dermatome in its proper position in storage aid
GB688R immediately after the operation.
! To avoid abrupt changes in temperature, use a single-chamber washer
disinfector.
! Select a cleaning and disinfecting cycle according to the following
criteria:
– Neutral detergent
– Cleaning temperature ≤ 50 °C
– Final
rinse with simultaneous thermal
demineralized water
Example: Miele Vario-TD process.
! Follow the instructions for use of the Aesculap ECCOS storage system
TA009721.

5.3 Care and inspection

! Allow the product to cool down to room temperature.
! Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
6

5.4 Sterilization method and parameters

Note
The carrier plates BA721, BA722 and BA723 are gamma-sterilized single-
use products. These products must not be steam-sterilized and are
intended for single use only.
The skin mesh dermatome can be sterilized in its storage aid.
! Sterilize the skin mesh dermatome immediately after cleaning.
! Store the skin mesh dermatome in a suitable wire basket (e.g. JF214R).
! Sterilize with steam, taking note of the following:
Carry out steam sterilization via an approved steam sterilization
procedure (e.g. sterilizer in conformance with EN 285/ANSI/AAMI/
ISO 11134-1993,
conformance with EN 554/ISO 13683). In the fractioned vacuum
process, the sterilization must be performed using the 134 °C/2 bar
program with a minimum holding time of 5 minutes.

6. Maintenance

To ensure reliable operation, Aesculap recommends that maintenance be
performed after 300 reprocessing cycles or at least once per year.
! For service, maintenance and repairs, please contact the B. Braun/
Aesculap representative in your country, see Technical Service.
disinfection with
ANSI/AAMI
ST46-1993
and
validated
in

Publicidad

Tabla de contenido
loading

Tabla de contenido