7. Troubleshooting list
Error
Finding
Cutting cylinder 2 fails to
Foreign body in device
rotate
–
Excessive axial slack of cutting
cylinder
Cutting cylinder 2 moves
Strong force required on hand ratchet
stiffly
12
Skin not cut through
Symbol on scale 18 shows scoring
completely
Symbol on scale 18 shows cutting;
incorrect fine adjustment
–
Cutting tips shiny, hand ratchet 12
stiff
Cutting tips broken off
8. Technical Service
For service, maintenance or repairs contact your B. Braun/Aesculap
representative.
Carrying out modifications on medical equipment renders the guarantee,
and any licenses, null and void.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 7461 95-2700
Fax:
+49 7461 16-2887
E-mail: ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
9. Accessories/Spare parts
Designation
Replacement cutting cylinder
Carrier plate factor 1.5
Carrier plate factor 3
Carrier plate factor 6
Cause
Cutting cylinder jammed
Hand ratchet 12 not engaged
Cutting cylinder loose
Defective bearing
Incorrect setting
Screws 11 loose
Cutting cylinder 2 worn
Cutting cylinder 2 defective
Designation
Hand ratchet
Basket storage aid
Storage aid
10. Technical data
Model
Dimensions (L x W x H)
Weight
Length of carrier plate
Width of carrier plate
11. Disposal
Art. no.
BA725R
The skin mesh dermatome BA720R is made of aluminum and stainless
BA721
steel.
! Dispose of the device with other recyclable metals.
BA722
BA723
Remedy
Remove foreign body
Engage hand ratchet 12
Engage cutting cylinder and secure with
lock 8
Repair by manufacturer
Turn the setting wheel so that symbol
"cutting" faces up; select the required fine
adjustment, see Safe operation
Select the required fine adjustment, see
Safe operation
Tighten screws by hand or with screwdriver
13
Replace cutting cylinder 2, see Safe
operation
Replace cutting cylinder 2, see Safe
operation
Art. no.
BA726R
BA727R
GB688R
BA720R
150 mm x 195 mm x 105 mm
4.600 g
220 mm
75 mm
7