2. Align the dot on the nosecone with the dot on
the neck of the handpiece.
3. Push the nosecone onto the handpiece so that
the dots are on top of each other.
4. Twist the nosecone clockwise until it locks into
place. The dot on the nosecone must now align
with the dot on the handpiece.
5. Remove the tip cleaner from the flue, and
slide the flue over the tip and onto the
nosecone. Markings on the flue indicate the tip
the flue should cover:
S
Standard Tip
Mi
MicroTip™
P
PrecisionTip™
M
MacroTip™
Make sure to push the flue base completely over
the nosecone base, and that the end of the flue
covers the pre-aspiration holes.
Attaching the Sterile Manifold Tubing:
If you receive the sterilized handpiece without
the manifold tubing already clipped onto the
handpiece cable, you must attach the manifold
tubing now. Follow this procedure.
6
1. Remove the manifold tubing and clips from the
packaging. Uncoil some of the tubing.
2. Snap the handpiece clip in place on the
handpiece housing.
3. Insert the irrigation tube from the flue into
the handpiece clip.
4. Connect the manifold irrigation tubing to the
Luer fitting on the handpiece flue.
5. Connect the manifold suction tubing and clip to
the suction port in the nosecone.
6. Push one to three manifold tubing clips onto the
handpiece cable.