PERFORMANCE SPECIFICATIONS:
The table below shows A
Masimo SET Oximetry Technology in a clinical study.
SpO
90-100%
80-90%
70-80%
70-100%
SaO
versus error (SpO
2
(Accuracy Root Mean Square) values measured using rainbow sensors with
RMS
70 - 100%
A
2
rms
1.57 (1,57) %
1.80 (1,80) %
2.47 (2,47) %
1.98 (1,98) %
– SaO
) with linear regression fit and upper 95% and lower 95% limits of agreement.
2
2
rainbow Series
SpO2
60-70%
70-80%
60-80%
rainbow Series
70 - 100%
60 - 80%
120
60 - 80%
A
rms
3.42 (3,42) %
2.49 (2,49) %
2.99 (2,99) %
9357F-eIFU-1117