• The suction machine is intended exclusively for the collection of NON-flammable fluids. It is not suitable
for use in the presence of a flammable anaesthetic mixture with air, or with oxygen or nitrous oxide.
• Always keep the power supply and the power cable away from hot surfaces.
• Keep the power supply cable away from animals (for example, rodents) which could damage the insula-
tion.
• Do not immerse the device in water; in the event of immersion disconnect the plug immediately. Do
not remove or touch the immersed device; unplug the power cable first. Immediately bring the device
to an authorised FLAEM service centre or to your trusted dealer.
• The device, power supply and battery casing are not waterproof. Do not wash the device under running
water or by immersion and keep it safe from being sprayed by water or other liquids.
• Do not expose the device to particularly extreme temperatures.
• Keep the device and power supply away from sources of heat, direct sunlight or excessively hot places.
• Repairs must be done only by authorised FLAEM personnel. Unauthorised repairs void the warranty and
may be hazardous for the user.
ATTENTION: Do not modify this device without authorisation from the manufacturer.
• The Manufacturer, the Vendor and the Importer shall be held responsible for safety, reliability and perfor-
mance only if: a) the device is used in compliance with the instructions for use b) the wiring where the
device is being used is in compliance with safety regulations and current laws.
• The Manufacturer must be contacted to communicate unexpected problems/events related to function-
ing.
• Ensure that the connection tubing and the canister lid have been carefully sealed in order to avoid loss of
suction.
• Do not upset the canister while the device is working as the liquid may be aspirated into the device, dam-
aging the pump. Should this occur, immediately switch off the suction machine, then empty and clean the
canister (send it to an authorised FLAEM service centre).
• When the safety device (10) intervenes, suction is halted. Switch off the device, empty the canister (9) and
carry out the cleaning procedure.
• We recommend personal use of accessories, collection canisters and connection tubing to prevent the
risk of contagious infection.
• The aspirator tip and the manual vacuum flow control are single use sterilised products and must be
replaced after each use.
• Check the expiration date on the original aspirator tip packaging and make sure the sterile packaging is
intact. FLAEM NUOVA declines any responsibility for harm to the patient correlated to the deterioration
of the aforementioned sterile packaging due to third party handling during the original packaging of the
entire device.
• The 12V DC power cable for vehicle power outlets is equipped with a safety fuse, which can be inspected
in the event of a fault.
• The power supply provided has been specifically designed for use with the Aspira Go series aspirators.
Never use the power supply with other devices or for uses other than that specified in this manual and
never use the Aspira Go series aspirators with other power supplies.
INSTRUCTIONS FOR USE
Before each use, the accessories must be carefully inspected to ensure the absence of dust, incrustation,
clots or liquid substances both inside the connecting tubing and in the canister and its respective lid. Further-
more, they must be cleaned following the instructions rigorously as stated in the "CLEANING, SANITISATION,
DISINFECTION, STERILISATION" paragraph. We recommend personal use of the accessories, the collection
canisters and the connecting tubing to prevent risk of contagious infection.
1. Operation with internal battery:
1.1. The device is supplied with the battery partially charged and it is recommended to charge it before use.
Insert the battery as shown on page 1. Release, lift and remove the lid of the battery compartment lo-
cated at the back of the device (dwg.1,2,3); insert the battery, fitting the connector properly (dwg.4,4a);
put the battery compartment lid back in place and close it (dwg.5,6). Recharge the battery by following
the instructions in section 2.3.1.
1.2. Switch the device on by pressing the switch (2).
1.3. At the end of each suction procedure, always turn the switch (2) to 0 (even if the battery is flat) to save
the battery life. If during use the red LED (1c) flashes and the buzzer sounds, it means that the device's
battery charge is about to run out, in this case recharge the device as soon as possible.
12