FLAEM Aspira Go P1611EM-20 Manual De Instrucciones página 16

Aspirador quirúrgico portátil con baterías
Ocultar thumbs Ver también para Aspira Go P1611EM-20:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 49
3.
Instructions for suction on the patient:
3.1.Connect the accessories, referring to the "CONNECTION DIAGRAM" on the cover.
ATTENTION: in addition to being antibacterial/antiviral, the hydrophobic filter also acts as secondary
protection device and stops any fluids that the primary safety device in the canister lid cannot manage
to stop.
Follow the directions below for proper filter assembly:
Wording:
Side to connect to the VACUUM
inlet (12) of the collection canister
3.2. Switch the device on by following the instructions for the required operation. Always use the device on a
flat surface; this will ensure the proper operation of the inflow safety device to prevent the inflow of fluid
in the suction device.
3.3. For more comfortable aspirations, set the desired vacuum value (bar) using the vacuum regulator (4).
Turn the knob clockwise to increase vacuum and turn it anticlockwise to decrease vacuum; these values
can be read on the "vacuum gauge" instrument (3).
3.4. Place your finger on the manual vacuum flow control (15) and, activating it in pulses, begin suction on
the patient using the aspirator tip.
3.5. Once the process is complete, switch off the device.
3.6. Empty and clean the canister and connecting tubes.
CLEANING, SANITISATION, DISINFECTION, STERILISATION
Switch off the device before any cleaning procedure and unplug the power cable from the socket.
DEVICE CLEANING
Use only a damp cloth with antibacterial soap (non-abrasive and with no solvents of any sort).
COLLECTION CANISTER AND CONNECTING TUBING
• Detach the aspirator tip (16), the manual vacuum flow control (15) and the tubing (14) from the canister (9).
Disconnect the tubing (6/8) from both the canister and the filter (7). Remove the canister from its holder,
keeping it upright, and empty it (in the WC at home, in the biological waste container in the hospital) and
clean it, disassembling it as shown below:
1. Remove the safety
device (.10)
from the canister lid
(.11)
Wording:
Aspirator side
Side to connect to the AIR INLET
of the suction machine
2. Remove the float (.10a)
from the safety device (.10)
14
3. Remove the
silicone gasket
(11a) from the
canister lid (.11)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Aspira go p1611em-30

Tabla de contenido