Directions for Use
Connecting the Precision Spectra
OMG Adapter
1. Inspect the Precision Spectra OMG Adapter,
OMG Connector-M, or OMG Connector-A for
damage. If damaged, do not proceed. Return
the damaged component(s) to Boston Scientific
Neuromodulation Corporation.
2. Turn off the Medtronic/St. Jude (Abbott) External
Trial Stimulator (ETS).
CAUTION
Failure to turn off the Medtronic/St. Jude
(Abbott) ETS before disconnecting the OR
Cable may result in excess stimulation.
3. Disconnect the proximal end(s) of the Medtronic/
St. Jude (Abbott) OR Cable Assembly from
the Boston Scientific ETS and insert into the
corresponding port on the OMG Connector:
OMG Connector-A
A.
Disconnect the proximal end(s) of the
St. Jude (Abbott) OR Cable Assembly
and insert it into the corresponding port
on the OMG Connector-A. Plug the St.
Jude (Abbott) OR Cables into the OMG
Connector-A socket(s) labeled "1-L" (left)
and "2-R" (right). If only a single lead is
being used, connect it to "1-L".
OMG Connector-M
B.
Medtronic Twist-Lock Screening Cables
(Model # 3550-03)
• If the trial is using the Twist-Lock
Screening Cables, plug the Twist-
Lock Screening Cable into the OMG
Connector socket(s) labeled "1-L"
(left) for a single-lead trial. The lead
shows up as contacts 1-4 in the Bionic
Navigator™ 3D software.
• For a two-lead trial, connect the
Medtronic OR Cables into "1-L" (left)
and "2-R" (right). The leads show up
as contacts 1-4 and 9-12 in the Bionic
Navigator 3D software.
Medtronic Snap-Lid Connector Cable
(Model # 355031) or Medtronic Multi-lead
Trialing Cable (Model # 355531).
• If the trial is using the Snap-Lid
Connector Cable or Multi-Lead
Trialing Cable, plug it into the OMG
Connector-M socket labeled "Single
Connector." If 2 four contact-leads are
placed in the Snap-Lid Connector Cable
or Multi-Lead Trialing Cable, the first
and second leads show up as contacts
1-4 and 5-8, respectively, in the Bionic
Navigator 3D software. The 4-contact
leads should be programmed in Manual
Mode.
• If 2 eight-contact leads are used, the
leads show up as contacts 1-8 and 9-16
in the Bionic Navigator 3D software.
4. Place the OMG Connector in a secure location
on a flat surface so that patient movement will
not cause the connector to fall.
5. Align and connect the "1-L" and "2-R" ports on
the Precision Spectra OMG Adapter and the
respective OMG Connector.
6. Making sure that the Boston Scientific ETS is
OFF, connect the Boston Scientific ETS to the
port labeled "CD" on the Precision Spectra OMG
Adapter.
CAUTION
Failure to turn off the Boston Scientific ETS
before connecting the Precision Spectra OMG
Adapter may result in excess stimulation.
Precision Spectra™ OMG™ Adapter
Directions for Use
92089844-02 3 of 81