Directions for Use
Connecting the Precision Spectra
OMG Adapter
1.  Inspect the Precision Spectra OMG Adapter, 
OMG Connector-M, or OMG Connector-A for 
damage.  If damaged, do not proceed. Return 
the damaged component(s) to Boston Scientific 
Neuromodulation Corporation.
2.  Turn off the Medtronic/St. Jude (Abbott) External 
Trial Stimulator (ETS).
CAUTION
Failure to turn off the Medtronic/St. Jude 
(Abbott) ETS before disconnecting the OR 
Cable may result in excess stimulation.
3.  Disconnect the proximal end(s) of the Medtronic/
St. Jude (Abbott) OR Cable Assembly from 
the Boston Scientific ETS and insert into the 
corresponding port on the OMG Connector:
OMG Connector-A
A. 
Disconnect the proximal end(s) of the 
St. Jude (Abbott) OR Cable Assembly 
and insert it into the corresponding port
on the OMG Connector-A. Plug the St. 
Jude (Abbott) OR Cables into the OMG 
Connector-A socket(s) labeled "1-L" (left) 
and "2-R" (right). If only a single lead is 
being used, connect it to "1-L".
OMG Connector-M
B. 
Medtronic Twist-Lock Screening Cables 
(Model # 3550-03)
• If the trial is using the Twist-Lock
Screening Cables, plug the Twist-
Lock Screening Cable into the OMG 
Connector socket(s) labeled "1-L" 
(left) for a single-lead trial. The lead 
shows up as contacts 1-4 in the Bionic
Navigator™ 3D software.
• For a two-lead trial, connect the 
Medtronic OR Cables into "1-L" (left) 
and "2-R" (right). The leads show up 
as contacts 1-4 and 9-12 in the Bionic
Navigator 3D software.
Medtronic Snap-Lid Connector Cable 
(Model # 355031) or Medtronic Multi-lead 
Trialing Cable (Model # 355531).
• If the trial is using the Snap-Lid
Connector Cable or Multi-Lead 
Trialing Cable, plug it into the OMG 
Connector-M socket labeled "Single 
Connector." If 2 four contact-leads are 
placed in the Snap-Lid Connector Cable 
or Multi-Lead Trialing Cable, the first 
and second leads show up as contacts
1-4 and 5-8, respectively, in the Bionic 
Navigator 3D software. The 4-contact 
leads should be programmed in Manual
Mode.
• If 2 eight-contact leads are used, the 
leads show up as contacts 1-8 and 9-16 
in the Bionic Navigator 3D software.
4.  Place the OMG Connector in a secure location 
on a flat surface so that patient movement will 
not cause the connector to fall.
5.  Align and connect the "1-L" and "2-R" ports on 
the Precision Spectra OMG Adapter and the
respective OMG Connector.
6.  Making sure that the Boston Scientific ETS is 
OFF, connect the Boston Scientific ETS to the 
port labeled "CD" on the Precision Spectra OMG 
Adapter.
CAUTION
Failure to turn off the Boston Scientific ETS 
before connecting the Precision Spectra OMG
Adapter may result in excess stimulation. 
Precision Spectra™ OMG™ Adapter
Directions for Use
92089844-02    3 of 81