Precision Spectra™ OMG
St. Jude OR Cable Assembly
OR Cable Assembly
St. Jude Lead
Trial lead, 4-Channel
Trial lead, 8-Channel
Axxess® Trial Lead
Octrode®
Quattrode®
Axxess®
Lamitrode® 22
Lamitrode® 4
Lamitrode® 8
Lamitrode® 44 and 44C
Lamitrode® Tripole 8 & 8c
Lamitrode® Tripole 16 & 16c
Lamitrode® 88 & 88c
Penta
Lamitrode S-8
Axxess, Octrode Quattrode and Lamitrode are trademarks of Advanced
Neuromodulation Systems
* These models numbers should be programmed in
manual mode
** Program only the mapped contacts on the
representative Boston Scientific lead
Precision Spectra™ OMG
90870349-03 REV C
Model Number
3009
Model Number
3046**
3086
4044**, 4054**
3183, 3186, 3189,
3181
3143**, 3146**,
3149**, 3141**,
3153**, 3156**,
3159**, 3151**
4143**, 4146**,
4153**, 4156**
3222*
3240**, 3254**,
3255**
3280
3244*,
3262*, 3263*,
3245*,3264*, 3265*
3208,* 3210*
3214, 3219
3288*, 3289*
3228*
3283, 3286, 3268,
3269
2 of 72
Directions for Use
Connecting the Precision Spectra OMG
1. Inspect the Precision Spectra OMG, OMG
Connector-M, or OMG Connector-A for
damage. If damaged, do not proceed. Return
the damaged component(s) to Boston Scientific
Neuromodulation or your healthcare professional.
2. Turn off the Non-Precision Spectra External Trial
Stimulator (ETS).
CAUTION
Failure to turn off the non-Precision Spectra
ETS before connecting the OMG Connector
may result in excess stimulation.
3. Disconnect the proximal end(s) of the non-
Precision Spectra OR Cable Assembly and insert
into the corresponding port on the OMG Connector:
A. OMG Connector-A
Disconnect the proximal end(s)
of the non-Precision Spectra OR
Cable Assembly and insert it into
the corresponding port on the OMG
Connector-A. Plug the non-Precision
Spectra OR Cable Assembly into the
OMG Connector-A socket(s) labeled "1-
L" (left) and "2-R" (right). If only a single
lead is being used, connect it to "1-L".
B. OMG Connector-M
Medtronic Twist-Lock Screening Cables
(Model # 3550-03)
• If the trial is using the Twist-
Lock Screening Cables, plug the
non-Precision Spectra OR Cable
Assembly into the OMG Connector
socket(s) labeled "1-L" (left) for a
single-lead trial. The lead shows
up as contacts 1-4 in the Bionic
Navigator™ 3D software.