Caution!
•
If the device does not behave as described in this in-
struction manual, it may not be used. Have the device
repaired immediately by the manufacturer Weinmann
or a dealer authorized by the manufacturer.
•
The MEDUCORE Easy is only guaranteed to be ready
for use if it has sustained no damage or misuse since
the last self-test.
•
Check at regular intervals that the device and its ac-
cessories are working properly (see "Performing the
function check" on page 37).
•
Do not deliver a shock if the defibrillation electrodes
are in contact with each other or are not in close con-
tact with the patient's bare chest.
•
Charging and delivering defibrillation energy may im-
pair the function of devices in the vicinity of the
MEDUCORE Easy. Perform a function check on these
devices before using them again.
Caution!
•
Electrical and magnetic fields may impair the function
of the MEDUCORE Easy. Keep the MEDUCORE Easy
two meters away from mobile telephones, radio
equipment, X-ray machines, fluorescent tubes etc.
•
Do not immerse either the MEDUCORE Easy or its ac-
cessories in liquids. No liquid may penetrate the
MEDUCORE Easy or its accessories. If this instruction is
not followed, the device and its accessories may be
damaged or user, patient and third parties put at risk.
•
If third-party items are used, functional failures may
occur and fitness for use may be restricted. Biocom-
patibility requirements may also not be met. Please
note that in such cases, any claim under warranty and
liability will be voided if neither the accessories nor
genuine replacement parts recommended in the in-
structions for use are used.
•
To prevent infection or bacterial contamination,
please observe section "Hygiene treatment" on
page 35.
Safety instructions
EN
15