Emc / Technical Description - Medela Swing Instrucciones De Uso

Ocultar thumbs Ver también para Swing:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 14

EMC / Technical Description

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2001)
IEC 60601-1-2, Table 201
Electromagnetic Emissions
The Swing is intended for use in the electromagnetic environment specified below.
The customer or the user of the Swing should assure that it is used in such environment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 60000-3-3
Warning – The use of accessories and cables other than those specified may result in increased
emissions or decreased immunity of the Swing breastpump.
The Swing breastpump should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the Swing breastpump should be observed to verify normal operation in the
configuration in which it will be used.
The Swing breastpump was found to have no essential performance by a risk analysis and thus, was
not tested for immunity to electro magnetic disturbances (IEC60601-1-2:2001, 6.8.3.201 i)
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The Swing uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
The Swing is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.

Publicidad

Tabla de contenido
loading

Tabla de contenido