13 Service and Maintenance continued
Technical Support
For technical support and product information for the 6000TOCi, contact your local
METTLER TOLEDO Sales Office or representative.
Once the filter replacement has been completed, the replacement must be registered with
the 6000TOCi to reset the filter operating information. At the M800, enter the TOC
Maintenance menu at Path: Calibrate (c)/TOC Maintenance c Replace Filter button to
register the filter replacement.
TO 6000TOCi
TO 6000TOCi
REPLACING THE FILTER ELEMENT FOR THE HIGH CAPACITY INLET FILTER
Draining and Shipping Instructions
The 6000TOCi contains a glass coil which facilitates oxidation of the water sample. Wa-
ter freezing within the coil during transportation will cause this coil to break. Therefore, it
is important that all water be removed from the TOC sensor prior to shipment.
To drain the water, connect the TOC sensor to AC power. To put the flow control valve into
the 'Drain' mode, press and hold the UV Lamp button for 5 seconds until the Sensor Sta-
tus LED begins flashing. This mode will fully open the automatic flow control valve to
allow draining the sensor more easily. Once the valve has been placed into the 'Drain'
mode, use the supplied plastic syringe to push the water out of the sensor, blowing from
the inlet, and pushing water out through the sensor outlet.
High Capacity Inlet Filter Replacement
The 6000TOCi Sensor includes a high capacity filter (shown
below) containing a filter element that should be replaced
(P/N 58 091 551, package of 2) after approximately 6
months of operation, depending on water quality conditions.
Detailed instructions to replace this filter are included in the
replacement package.
PLACE 3/4" WRENCHES HERE
FILTER HOUSING
FILTER
(B)
UNSCREW INLET NUT
INLET SIDE OF HOUSING
COMPRESSION NUT (A)
OPEN
END
SPRING
15