1.1.3. Remove the mug and throw away the muddy water. Repeat this proce-
dure several times till clean water comes out from the Analyzer.
1.1.4. Done.
1.2. Cleaning out at the end of a working day.
This procedure prevents formation and collection of fat and "Milk stone" de-
posits into the sensor. Milk stone consists of milk solids, calcium, magne-
sium, iron, sulfates, etc. Milk and water mineral deposits become hardened
and layered on the sensor and vinyl pipes inner surfaces, which contact with
milk and disturbs the milk analyzer work. Cleaning will be effective if a re-
agent that attacks the "milk stone" is used. We recommend 2% solution of
the alkaline cleaner EkoDay to be used as a daily cleaning solution.
Preparation of 2% EkoDay working solution:
a.Use pipette to add 10 mL EkoDay to glassware with 490 mL distilled water.
b.Pour into a labeled container.
Take care this solution does not contact your eyes or skin!
Attention! Use only 2% EkoDay working solution as a cleaning agent. The
EkoDay concentrate can damage your analyzer!
Steps:
1.2.1. First clean the analyzer with pure water as it is described in proce-
dure 1.1. (see above).
Attention! Using 2% solution of the alkaline cleaner EkoDay without first
cleaning analyzer for removing the fats and proteins will result in fixing the
"milk stone" to the surface. Always clean analyzer with pure water before
using the cleaning agent.
1.2.2. Fill the measuring mug with 2 % solution of the alkaline cleaner EkoDay
(25°- 40°C), put the analyzer in cleaning mode, set 20 cycles and press the
OK button. When the CLEANING stage is over CLEANING END is shown on
the display.
1.2.3. Fill the measuring mug with clean water, put the analyzer in cleaning
mode, set 5 cycles and press the OK button. When this procedure is done,
remove the mug and throw away the water. Fill the measuring mug with clean
water and repeat this procedure 3-4 times.
1.2.4. Done
2. PERIODICAL PROCEDURE
To ensure a good work of the Milk Analyzer it is advisory to clean the device at
least once a week strictly performing underwritten procedure.
This procedure uses 5% solution of acid cleaner EkoWeek as a periodical
cleaning solution.