TECH TIPS
Use 30mm screws for attaching to a thick check socket and the 20mm for a thinner check socket
▪
Recommended amount of screw length projecting out of the socket should be between .450-.550˝
Mounting surface must be flat and parallel in order for lock to operate properly.
▪
Do not remove the backing plate on the shoulder joint.
▪
The 1 /16" hex-head screw, accessible through the center of the joint can be removed and relocated to
▪
change the position of the center stop. There are 6 possible locations, 60° apart.
With the bent plate, the user can internally rotate 15° toward the body mid-line when the upper arm is
▪
straight ahead. If you prefer a straight line, then cut off the bent end of the Humeral Plate and mount the
straight end to the Yoke.
CAUTIONS AND WARNINGS
LOAD CAPACITY MAY NOT EXCEED 50LBS (22.7 KG) INCLUDING PROSTHESIS.
WARRANTY INSPECTION / MAINTENANCE INFORMATION
INSPECTION GUIDELINES
Visual/Audible signs of wear, cracks, damage, etc.
▪
Ensure that the joint moves freely without obstruction.
▪
Ensure that the lock engages and disengages properly.
▪
TECHNICAL ASSISTANCE / EMERGENCY SERVICE 24/7/365
College Park's regular office hours are Monday through Friday, 8:30 am – 5:30 pm (EST). After hours, an
emergency Technical Service number is available to contact a College Park representative.
LIABILITY
The manufacturer is not liable for damage caused by component combinations that were not authorized
by the manufacturer.
CAUTION
College Park products and components are designed and tested according to the applicable official standards
or an in-house defined standard when no official standard applies. Compatibility and compliance with these
standard is achieved only when College Park products are used with other recommended College Park
components. This product has been designed and tested based on single patient usage. This device should NOT
be used by multiple patients. If any problems occur with the use of this product, immediately contact your
medical professional.