8. Check the applicator tip and, where provided, the electrode.
If contamination is still visible on the instrument, repeat pre-
cleaning steps 1 to 4.
9. If available: Retract the electrode for the following process-
ing steps.
9.6
Manual cleaning and disinfection
Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have the followoing pH val-
ues:
• pH value between 5.5 and 11 for: 20132-033, 20132-040,
20132-140
• pH value between 5.5 and 12.3 for: 20132-031, 20132-032,
20132-034, 20132-044, 20132-054, 20132-056, 20132-
057, 20132-084, 20132-133, 20132-211
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
1. Prepare a cleaning bath and a separate disinfectant bath.
2. Immerse the product completely in a cleaning bath, ensuring
that the product does not come into contact with other com-
ponents in the bath.
3. Thoroughly clean the surfaces with a soft brush/soft dispos-
able cloth.
4. Extend and retract the movable parts (at least 5 times).
5. Completely fill a disposable syringe with cleaning solution
and rinse the lumina (at least 5 times).
32