Cleaning, Disinfection, Sterilization; Safety Instructions For Processing - Erbe 20132-200 Manual Del Usuario

Vio apc-mangos
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58
OBJ_DOKU-188704-009.fm Seite 30 Donnerstag, 25. März 2021 11:48 11
2. Press the blue COAG button.
– For APC 300 + ICC: The respective COAG mode must be
activated on APC 300 and ICC. If this is not the case, the
handle is damaged.
– For APC 2 + VIO: The COAG mode must be activated on
the VIO. If this is not the case the handle is damaged.
3. Testing the ReMode button
setting options on the VIO. Press the ReMode button several
times. Each time, it must be switched back and forth be-
tween both setting options of the program on the VIO. If this
is not the case, the handle is damaged.
If an error message appears on the VIO electrosurgical unit or on
the ICC / APC 300 when plugging in the handle without a button
being pressed, the handle is defective.
If damaged, do not use this product!
7

Cleaning, disinfection, sterilization

7.1

Safety instructions for processing

Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
1.) Can only be tested with handle 20132-200 and connection of the
handle on an APC 2 in combination with a VIO D unit.
1)
: Switch on a program with two
30

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

20132-201

Tabla de contenido