5.2
Selecting a suitable HF electrode
This handle can also be operated with Erbe HF electrodes (shaft
diameter: 4 mm).
Erbe recommends the use of Erbe HF electrodes. If the handle is
operated with HF electrodes from another manufacturer, please
observe the following: The joint from handle to HF electrodes of
other manufacturers is not standardized. The HF electrode of an-
other manufacture, although it may appear to fit the Erbe han-
dle, may have different dimensions and tolerances, a different
type of insulation, etc. Such differences hold risks for patient and
staff. Erbe Elektromedizin is glad to provide more information as
required.
5.3
General information about protection of the user,
patient, and product
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Erbe Elektromedizin recommends performing an electrical con-
tinuity test before every use of this product.
If the cable conductor is broken or the electrical transmission of
the cable is interrupted in any other way, this may lead to arc
formation in the patient return line or in the active circuit; this
may cause burns to the patient/surgical staff or start a fire.
Checks other than those described in these instructions for use
do not need to be carried out.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
If damaged, do not use this product!