4
How to use
4.1
General
Always open the gas cylinder valve slowly and carefully.
Always close the gas cylinder valve after use.
Do not use any kind of tool to operate, attach or detach the pres-
sure reducer.
Do not smoke in the vicinity of gas cylinders and pressure reduc-
ers.
4.2
Changing the cylinder
Also observe the illustrations at the start of these instructions.
1. Close the gas cylinder valve! The gas cylinder valve may be
a little difficult to turn in some cases.
2. Remove the pressure line (quick coupling) from the connec-
tion on the rear of the APC device.
3. The pressure line still contains a residual amount of argon
that must be discharged. To do so, place the quick coupling
onto the purge post on the rear of the APC device and press
against it. The residual argon is now discharged with a loud
hissing sound (check: 0 must be shown on the high pressure
manometer).
4. Remove the pressure sensor from the connection on the rear
of the APC unit.
5. In the counter-clockwise direction: turn and remove the re-
tainer nut of the pressure reducer by hand from the argon
gas cylinder (do not use any tools).
6. Change the argon gas cylinder.
22