OBJ_DOKU-27191-012.fm Seite 25 Mittwoch, 2. Dezember 2020 12:17 12
6.5

Conducting a performance test

1. Moisten a gauze strip with a sterile NaCl solution.
2. Bipolar LAP forceps: Grasp the gauze strip with the jaws and
activate the HF current.
Bipolar LAP scissors: Press the open jaws onto the gauze
strip and activate the HF current.
Steam will form after approx. 2 to 3 seconds if the instrument is
working properly.
If damaged, do not use this product!
6.6

Using the product

The instrument is ready for use with 5 mm trocars. To insert into
a trocar sleeve, the instrument jaws must be closed.
To close the jaws, press the handle together. To open the jaws,
pull the handle apart.
ERGO handle (20195-245): The upper part of the handle can be
rotated freely relative to the lower part. The required rotational
position can be set using the locking knob.
Do not coagulate unless the contact surfaces are within your
field of vision. Do not touch any metallic instruments during this
process.
LAP scissors (20195-242): The slower the cut is made (with
COAG mode activated), the larger the coagulation zone at the
edge of the cut.
6.7

Precleaning the product in the procedure room

1. Use a soft cloth (such as a compress) to remove any obvious
contamination immediately after use and/or place the prod-
uct in deionized water.
25